Cazacu AC, Chung SE, Greer J, Demmler GJ. Comparison of the directigen flu A+B membrane enzyme immunoassay with viral culture for rapid detection of influenza A and B viruses in respiratory specimens. J Clin Microbiol. 2004 Aug;42(8):3707-10
submited by kickingbird at Aug, 30, 2004 9:59 AM from J Clin Microbiol. 2004 Aug;42(8):3707-10

Comparison of the directigen flu A+B membrane enzyme immunoassay with viral culture for rapid detection of influenza A and B viruses in respiratory specimens.

Cazacu AC, Chung SE, Greer J, Demmler GJ.


Department of Pediatrics, Baylor College of Medicine, Houston, Texas 77030, USA.

The performance of a commercially available, rapid membrane enzyme immunoassay for influenza A and B virus detection was compared to that of viral culture in 4,092 respiratory specimens collected from patients presenting with respiratory symptoms during the 2002-2003 influenza season. The test´s overall sensitivity was 43.83%, lower than previously reported but similar for detection of both influenza A and B viruses (42.98 versus 44.76%). However, specificity, 99.74%, was excellent for both influenza A and B viruses (99.82 versus 99.92%). These values make this test a very good confirmatory test when clinical suspicion is high, but a less accurate screening test for large populations.