Air sampling and simultaneous detection of airborne influenza virus via gold nanorod-based plasmonic PCR

Reliable and sensitive virus detection is essential to prevent airborne virus transmission. The polymerase chain reaction (PCR) is one of the most compelling and effective diagnostic techniques for detecting airborne pathogens. However, most PCR diagnostics rely on thermocycling, which involves a time-consuming Peltier block heating methodology. Plasmonic PCR is based on light-driven photothermal heating of plasmonic nanostructures to address the key drawbacks of traditional PCR. This study introduces a methodology for plasmonic PCR detection of air-sampled influenza virus (H1N1). An electrostatic air sampler was used to collect the aerosolized virus in a carrier liquid for 10 min. Simultaneously, the viruses collected in the liquid were transferred to a tube containing gold (Au) nanorods (aspect ratio = 3.6). H1N1 viruses were detected in 12 min, which is the total time required for reverse transcription, fast thermocycling via plasmonic heating through gold nanorods, and in situ fluorescence detection. This methodology showed a limit of detection of three RNA copies/μL liquid for H1N1 influenza virus, which is comparable to that of commercially available PCR devices. This methodology can be used for the rapid and precise identification of pathogens on-site, while significantly reducing the time required for monitoring airborne viruses.